S-Parameter Conversion

S-parameter conversion function allows for conversion of measurement results (Sab) to the following parameters:

Parameter Equation
Impedance in reflection measurement (Zr)
Admittance in reflection measurement (Yr)
Impedance in transmission measurement (Zt)
Admittance in transmission measurement (Yt)
Inverse S-parameter
Equivalent admittance in transmission shunt measurements (Ytsh)
Equivalent impedance in transmission shunt measurements (Ztsh)
S-parameter complex conjugate
Z0a – characteristic impedance of Port a,
Note Equations for Zr, Zt, Yr, Yt are approximate. The general method of converting S - parameters to Z, Y, H, T, ABCD - parameters is presented in the next section. The reason for using the approximate method is the measurement speed, as only one S – parameter is used in the calculations, whereas the general method requires measurement of the full matrix of S-parameters.

The S-parameter conversion function can be applied to an individual trace of a channel. Before enabling the function, first activate the trace.

To enable/disable the conversion, use the following softkeys:
To select the conversion type, use the following softkeys:
Note All conversion types are indicated in the trace status field, when enabled.